Antares
Analytical Technical Bulletin for October 2001
Subject: Poor Analytical Results for P, C, S,
B, Se, As
Relevant Models: FSQ, DV, FoundryMate with 1/2 meter
vacuum option - PST / PS1-6 / PS2000 Plasma series
These problems can also occur
in the MOA, which will be written about in a later Bulletin
Relevant Matrix: Fe base (All), Non-Ferrous (primarily Pb,
Sn, Cu alloys)
Symptoms:
- Poor repeatability on above elements but other
major elements (for example: Ni, Cr, Mo in Fe Base .... Sb, Bi, in
Pb / Sn) have good repeatability
- Poor Standardization "N" factor for the
elements listed in "Subject" (N factor below 0.2 or above 2.00)
- Low intensity ratio even though high voltage setting is
near or at maximum (9 on RS2 / RS3 / CMU equipped units or 7 on MC20 equipped
units)
Most Probable Reason:
- Loss of far UV (ultraviolet) light
transmission for wavelengths below 2000 Angstrom. All the above elements have
wavelengths below 2000 A and these wavelengths require special circumstances
to transmit from the analytical "spark" in the sample stand to the
PMT detectors in the optical cell. Normal atmosphere will absorb these wavelengths
below 2000 A so either vacuum or argon are required for transmission.
Since normal glass will absorb ALL UV wavelengths below about 4000 A, all
the optics in your Baird spectrometer are either quartz, MgFl (magnesium fluoride)
or a MgFl / AlO2 coated diffraction grating.
Possible Causes of UV Loss:
- Dirty or discolored sample stand slide window: Solution: Clean
slide window with suitable lens cleaner. Inspect the center portion of the
window for a small brown discolored area about 10 to 15 mm in diameter. This
brown discoloration will be caused by strong ultraviolet light "burning"
the window and this UV burn can absorb the wavelengths below 2000 A. Sometimes
this "UV burn" can be cleaned using 100% pure Acetone or Triclorethene.
If Acetone or Triclorethene doesn't remove this UV burn then the window must
be replaced. WARNING: Lens cleaner or other cleaner MUST
NOT be used on the DV6-SN (Nitrogen) units. The DV6-SN uses
MgFl windows and optics and must be cleaned by a special method. Please contact
Antares Analytical for the correct procedure. This MgFl cleaning procedure
is not in the DV6 Manual.
- No argon purge of the argon purged entrance optics.
Solution: This is the entrance optics chamber located directly
behind the sample stand where the Hg bulb (FSQ, DV2-5 and all Plasma Series)
used for optical alignment is located. The DV6 / DV6E / FoundryMate do not
use a Hg bulb for optical alignment. Under the optical chamber, behind the
sample stand on all models, will be a small brass fitting connecting plastic
tubing to the optical chamber. This plastic line brings a constant stream
of argon from the gas panel to the chamber to eliminate air, because
air also absorbs UV wavelengths below 2000A. There should be a
small amount of argon pressure coming through this line. If there is no argon
then adjust the constant flow setting on the gas panel. Use the Baird
supplied manuals or contact Antares Analytical for advice.
- Dirty or UV burnt vacuum window: Solution: Like the slide window, the vacuum
window can eventually become dirty or UV burnt. Cleaning is done
using the same method as for the slide window. This window is located
on the vacuum cell door on the DV2 to 6 and all Plasma models. Bringing
the unit to atmosphere and opening the cell door allows this window to be
removed for cleaning. On the FSQ changing or cleaning this
window should be done by a qualified Service Engineer only. On the FSQ removing
& replainv this window involves removing the sample stand for access. The
sample stand must be re-aligned after this window is changed and if this
alignment isnot done correctly, your FSQ will not work correctly. Consult
your Baird supplied manuals or contact us for advice before cleaning
/ changing this vacuum window.
- UV Burnt Entrance Mirror: (Applicable to the FSQ, DV2-5,
all Plasma series only) Solution: The entrance mirror, located inside
the vacuum cell on all models, can not be cleaned and can only be replaced.
This must be done by a qualified Service Engineer during a normal Maintenance
visit. Replacement of this mirror requires precise re-alignment of the entrance
optics and entrance mirror for correct miror replacement. Please contact Antares
Analytical for more information. Mirror replacement and re-alignment is not
in the Baird supplied manuals.
- UV Burnt or oil covered grating: Solution: The grating is the heart of the
spectrometer's optical system. The grating can
almost never be cleaned and NEVER should the grating be touched
with ANYTHING as damage will occur. There are very certain times when a grating
can be cleaned but ONLY with very special cleaning solution, using a very
precise method and must be done only by a Qualified
Service Engineer. Gratings on an FSQ or MOA can re replaced at
the customer site and only by a trained Service Engineer. Antares Service
Engineers do have this training and experience. The DV and Plasma gratings can
not be replaced. On the FSQ sometimes oil from the vacuum
pump can blow back into the optical system and deposit on the grating. This
is usually caused by operator error when starting the spectrometer after a
period of time it has been shut down. Incorrect shutdown and re-starting of
the FSQ is almost always the main reason.
This monthly Technical Service
Bulletin has been sent to you by the Technical Service Staff of Antares Analytical.
Our purpose for sending this is to help the Baird spectrometer user get
the lowest cost, best use of your Baird spectrometer. If you have
any comments, suggestions for future Technical Bulletins, questions about the
month's topics or questions / problems about your Baird spectrometer please
contact us at:
bulletin@antaresanalytical.com
If you wish to be removed
from our email list please reply to bulletin@antaresanalytical.com
with REMOVE in the subject line.