Antares Analytical Technical
Bulletin for June 2002
Subject: Photomultiplier
Tubes (PMT)
Relevant Models: All
Baird Spectrometers
Relevant Matrix:
All Matrices
Background:
-
All Baird spectrometers
use PMT for element detection and to send the correct signal level to the
spectrometer electronics for correct percent calculation. Each element
channel will have one PMT dedicated to only that element.
-
A PMT is a photonic (optical / electronic)
device which detects the element light level coming from the grating
/ exit slit, and accurately sends an electrical current to the measuring
electronics.
-
A PMT is
a transducer (a component which changes one type of energy
into another ie: light energy into electrical energy) type analog device. Since all PMTs are analog devices,
this "PMT signal" is measured and converted, through an analog to digital (A/D)
converter, into a number the computer (a digital device) can use for
accurate element concentration reporting.
-
A faulty
or failed PMT can and will generate an inaccurate electrical signal
which will cause the computer to report an incorrect element concentration.
The good thing is that PMTs are a very stable, reliable device and many
customers will probably never see a problem. With average use the life expectancy
of the average PMT is measured in years, though with time they will
weaken, but again this is usually over years.
PMT Characteristics: The following general
characteristics are basic to all PMTs used in a Baird Spectrometer:
-
All
PMTs use a certain voltage from the high voltage power supply to work.
-
This
voltage is normally within the -450 to -1000 VDC range and the correct voltage
setting depends on the PMT sensitivity (discussed below), the sensitivity
of the element line being measured, and the relevant matrix. This is normally
factory selected and can be adjusted by the customer using thumbwheels,
or from the computer for spectrometers using an MC20. The exception
is for the old Baird spectrometers using the RS3, RS4 or RS6
readouts (DV3 .. certain DV5 and certain ICP models) which use a plug-in
resistor board for HV adjustment.
-
The
newest DV6E and OneSpark also use this plug-in resistor type adjustment
board.
-
On
some other type spectrometers (Labtest for example) this high voltage
is controlled by adjustable "attenuators" but the idea is the same.
-
All
PMTs have a spectral response outside the wavelengths listed below, but
for optimum response, should only be used for the ranges listed.
-
Electrical current is "induced" when element light hits
the the PMT "grid." This grid is the part of the PMT
which looks like a window screen and is generally made of platinum or rhodium.
It is this induced current which is sent down to the measuring electronics /
computer where it is measured and a percent concentration is calculated.
-
There
is a direct relationship between the amount of element in a sample and the
amount of light which hits the grid of the PMT and the amount of current
sent to the electronics. More sample = more light (photons)
= more current at a fixed high voltage.
-
PMTs
of a certain type (R1414 for example, but the same criteria for all PMTs) have
different sensitivities based upon the anode sensitivity
of the individual tubes. Generally the more anode sensitivity the more light
the PMT can detect. this is very important when analyzing certain very low
element concentrations. Examples would be Boron in Steel or Phosphorous
in Aluminum. For higher concentrations (examples are Nickel is Stainless
Steel or Zinc in Gunmetal) or when the line is used as a reference line
you want a lower sensitivity line.
-
Tubes
were graded by sensitivity by Baird and marked with either a number or color.
Some tubes (R928, R889, 1P21, R166 among several others) were not given
a sensitivity rating. Please see "Sensitivity Ratings" below for more details.
-
When
your Baird spectrometer was configured at the Baird factory (closed
by Thermo Electron in 2000) all these factors were taken into account
when choosing exactly which tubes for which lines in your particular spectrometer.
This is why you should replace a PMT with a closely equivalent PMT, however
any PMT of the same type will work but adjustments in high voltage setting
or optical attenuation may be necessary.
PMT Selection Criteria: The correct
PMT's are selected based on the element wavelength being measured and the relevant
matrix. The Baird spectrometers are designed to work in the 170 nm to 800
nm (1700 to 8000 Angstroms) range. There is an exception for the DV6S/N which
has been configured to measure Nitrogen, Oxygen and Hydrogen in the > 160
nm range but this is a special case requiring special MgFl optics, grating and
PMT.
1. Element Wavelength: Generally PMT's
are selected for the wavelength being measured along these general criteria.
This will be the same criteria for all Baird spectrometers, regardless of type.
-
170
nm to 240 nm (approximate) will use an R427, R1657, or R166
-
240
nm to about 440 nm which will use an R300, R1414 or IP28
-
440
nm to 800 nm which will usually use an R889, R928 or 1P21
-
Some
element lines of an element (example: Fe 259.9 nm which is used in
the .001 to 2.50% range), will be much more sensitive than others of the
same element (example: Fe 271.4 nm which is used as a reference line in
Ferrous matrix). In the this case
2. Matrix: Generally the specific Matrix
will determine which analytical lines will be used. The choice of tubes is dependent
on the wavelength and concentration range for the desired element.
3. Baird Spectrometer Model: Generally
the older the model spectrometer HA10 series, FAS Series and early DV2-3 series
will use only larger type tubes (R166, R928, 1P21 and 1P28). Later models of
the DV series (4 and 6), all ICP series, and HA12 will use a combination of
large and small (R1414, R1657, R300, R427, R889) type PMT. DV5,MOA, FSQ and
all mobile spectrometers (MAS MAXII, MS3) will use only the small PMT.
4. PMT Sensitivity Grading: PMT sensitivity
is graded according to the anode sensitivity and varies within tube type. This
sensitivity is NOT depend on an element wavelength: Generally as follow:
-
1P21, R166, R928, R889 - no grading
-
1P28 originally with a -1 to -5 with -1 being least
sensitive and -5 being most sensitive.
-
R427, R300: only graded with a standard sensitivity
or a -1 sensitivity for a more sensitive tube:
-
R1657, R1414: 2 grading systems were used over time
for these tubes. a 5 color system replaced with an 11 color system being
based on the Intl standard color coding for resistors with black being least
sensitive to white / white being most sensitive.
5. PMT Replacement Availability:
-
1P21,
R166, R928, R1259 are still available.
-
R427,
R300 are no longer available on the open market. Antares Analytical does
have a good supply in stock for customers needing a replacement
-
R1414,
R1657, R889 are no longer available. These have been replaced by a
6350 to 6358 series, again dependent on wavelength and sensitivity. They
are direct replacement for the R1414, R889 and R1657. Please contact Antares
Tech Support for advice & availability of these replacement PMT.
-
1P28
are still available but only in the 1P28-1 and 1P28-5 versions.
For further information or access to past monthly Technical Bulletins covering
a variety of subjects, please visit our website (under development) at www.antaresanalytical.com
For more information, to be added
to this list, or help with other subjects including Tech Support, spare
& replacement parts, or on-site Service by qualified Baird trained Antares
Analytical Service Engineers, please contact us at info@antaresanalytical.com
To
be removed from our email list, please reply with REMOVE in the subject heading
and we will never bother you again unless you request.
This email is not Spam and is intended
to give Baird spectrometers users a better understanding of their Baird spectrometer
and help them get best results from your spectrometer at the lowest reasonable
cost. We would like very much your comments, suggestions for future articles
and complaints. Antares Analytical needs you, the customer, to tell us how
we can better service you.