Antares
Analytical Bulletin for July 2002
Subject:
Loss of High Voltage (also called HV)
Relevant Units: All FSQ, all DV with MC20 and all ICP with MC20 readout.
Certain HA10-12 and Labtest spectrometers with MC20 upgrade will be noted where
appropriate. This does not cover the DV6E or Foundrymate which use a different
power supply and interlock system.
Relevant Matrix: Not Applicable
Symptoms:
- No HG signal on panel mounted
HG meter, no Hg reading on the computer screen during optical alignment (later
FSQ models without a panel mounted vacuum meter), no element signal while
aligning (number is less than 10) … DV6 only which uses an element line, not
an Hg bulb for optical alignment
- BAIRDAS program shows red in HV
box of the Routine Analysis screen of the program
- Sudden inability to get accurate
analytical results. See April 2002 Bulletin for more information on this.
- HV LED on the MC20 front panel
is out
Background Information
- Do not automatically assume because
there is no high voltage, the high voltage supply is bad. There can be several
possible reasons for no high voltage.
- Almost all FSQ / DV, ICP series units are vacuum units, but there are a
limited number of airpath (no vacuum pump) DV / FSQ units installed at various
customer sites. All HA10/12 units are air path.
- Where air units are different
from vacuum units, the differences will be explained, otherwise all information
will be applicable to vacuum and airpath units. Except where information is
applicable to all types, most of the information will be used only as background
for the HA10-12 units.
- The high voltage power supply,
located in the left front slot of the MC20, is interlocked in several different
ways, depending on the spectrometer. Please verify which spectrometer configuration
you have (Vacuum DV, airpath FSQ etc) and find the correct selection listed
later in this bulletin.
- The high voltage in interlocked
primarily to protect the PMTs (photomultiplier tubes – please see June 2002
Antares Tech Bulletin) from damage in case of loss of vacuum in the optical
cell. This is not a concern in the air path units.
- Another reason for the high voltage
interlock is optical. As the unit pumps down, the vacuum cell “stresses” and
optical dispersion shifts. A further explanation of this “dispersion shift”
isn’t necessary here. This “dispersion shift” is calculated into the spectrometer
design and for correct analytical results, the vacuum spectrometer must be
operated at vacuum. This is not a concern for air path units.
Basic High Voltage System:
- The High Voltage Power Supply is a step up transformer which takes 24 VAC
from the Torroidal transformer, steps it up to 1000 VAC, rectifies / inverts
to –1000 VDC, filters and exports to the Motherboard of the MC20.
- This 24 VAC supply voltage is run through the set of contacts on the K1
relay located on the high voltage power supply.
- The K1 relay is energized, and
contacts closed, when the correct interlock signal is received from the spectrometer.
- The 24 VDC for the interlock circuit
is generated within the MC20 and sent to the spectrometer interface board.
Basic Interlock Systems
by Type
- The DV2 to 5 series, all ICP Series,
and earlier FSQ with a front panel mounted vacuum gauge: The high voltage
will be interlocked through the contacts on the back of the vacuum gauge.
The on / off of the high voltage is controlled by the vacuum pressure in the
cell letting (or not letting) the red indicator needle fall below the 2nd
black HV set point needle. The set point needle on the vacuum gauge which
indicates vacuum pressure, opens and closes contacts within the vacuum gauge.
- DV6 / FSQ with no vacuum gauge:
The high voltage is interlocked through the electronic set point on the vacuum
thermocouple. This is a factory adjusted setting. When vacuum pressure in
the cell is low enough, an electrical control signal is sent down to the high
voltage power supply to turn on.
- DV / FSQ Air Path units: For these units the vacuum interlock
is wired permanently closed and high voltage will be present at all times
unless the MC20 or high voltage power supply are turned off.
- HA10 / 12: The high voltage interlock are 2 switches, wired in series, which
close as soon as the top optical cover is put in place and fastened down correctly
and tightly.
Possible Causes and Solutions:
High Vacuum Pressure (> 100 microns) or no Vacuum:
- Check to be sure your vacuum level
(red needle on the vacuum gauge) is below the 2nd (black) set point
on the older knob type vacuum gauges.
- Push the “high set point” button on the newer Pirani type vacuum gauges.
This set point should be about 50 and the red indicator needle should be below
it.
- Solution: Check
for correct operation of the vacuum system. Be sure the vacuum pump is operating
and the gate valve is open. More will be covered on the DV / FSQ / ICP vacuum
systems in the Antares Tech Bulletin in August.
Bad Thermocouple: The
thermocouple may not be putting out voltage to the high voltage power supply
interlock (FSQ & DV6 without a vacuum meter). Thermocouple may not be putting
correct reading to the vacuum gauge (front panel mounted vacuum meters … ICP,
DV2-5 and FSQ). Please contact Antares Analytical for further help in testing.
Bad Vacuum Meter: The spectrometers which use a front panel mounted vacuum meter
have the HV interlock signal going through one set of set points inside the
meter. This set point is manually adjusted from the front meter and is normally
the higher of the 2 black needles. Inside the gauge are a set of mechanical
contacts which are closed when the vacuum indicator needle moves below the 2nd
set point, closing the HV interlock circuit. It is common for these contacts
not to close even though the red indicator needle is lower in pressure than
the HV interlock set point. This is a common problem on some older units (DV4
and early FSQ for example). This is not repairable and contact Anatres Analytical
for advice.
Bad HV Power Supply: There are several things to check on the HV power supply
·
The K1 relay
has been a common failure in the past. If the relay is suspected, replace from
your stock and try again.
·
There are 6
fuses on the HV power supply. Check to be sure all are good and replace any
bad ones, with new fuses from stock.
·
Further than
this please contact Antares Analytical for troubleshooting the HV Power Supply
Board.
Bad Low Voltage Power Supply
(MC20): Please contact Anatres for further
advice
Bad Torroidal Transformer
(MC20): Please contact Antares for further
advice.
______________________________________________________________________
This Technical Bulletin is written
by Gerry Smith, Technical Services Manager of Antares Analytical. Having spent
22 years with Baird in various capacities, including many years as Final Test
/ Calibration Manager and Senior Service Engineer, International Operations,
he is well qualified to answer any Technical or Applications problems you may
have.
For further information or access to
past monthly Technical Bulletins covering a variety of subjects, please visit
our website (under development) at www.antaresanalytical.com
For more information, to be added to this list, or help with other subjects
including Tech Support, spare & replacement parts, or on-site Service by
qualified Baird trained Antares Analytical Service Engineers, please contact
us at info@antaresanalytical.com
To be removed from our email list,
please reply with REMOVE in the subject heading and we will never bother you
again unless you request.
This email is not Spam and is intended
to give Baird spectrometers users a better understanding of their Baird spectrometer
and help them get best results from your spectrometer at the lowest reasonable
cost. We would like very much your comments, suggestions for future articles
and complaints. Antares Analytical needs you, the customer, to tell us how we
can better service you.