Antares Analytical Tech Bulletin
April, 2003
Problem: No Burning in the MOA / FAS Sample Stand This Tech Bulletin
will deal only with the most common cause of this problem: A faulty “Source
On” Interlock System.
Relevant Models: MOA / FAS (all models: MOA-C …MOA-GT …. MOA- C/GT …Mil
MOA), HA10/12 configured as Oil Analysis unit, older DV2 models with the Oil
Analysis Module as a “2nd stand. Information about this problem,
as it relates to Baird metal spectrometers (DV, FSQ etc) can be found in the
November 2001 Antares Technical Bulletin.
Relevant Matrices: Oil and Fuel Analysis
Background Information: The MOA source is designed to give constant,
controlled & regulated, high voltage DC power to the analytical gap for
sample excitation. The source circuitry is a simple, well designed and rugged
package, manufactured to US DOD Mil-Spec qualification, and will provide years
of trouble free operation, when maintained and operated correctly. The MOA source
/ stand come in several configurations depending on the type sample being analyzed.
This Tech Bulletin will deal only with the MOA / FAS systems. Users of HA10
/ 12 or DV2 with an oil stand should contact us directly for further information.
System Interlock Failure: As said above, the most common cause of no
burning in the sample stand is a defective interlock switch or failure of the
interlock circuitry. The function of this interlock system is to insure the
door is closed burning sample analysis because:
- The sample burning puts out a large amount of bright UV (ultraviolet) light
which can be damaging to the eyes.
- The high voltage across the analytical electrodes is approximately 22,000
VDC @ 2-4 amps and can be fatal.
- The high voltage arc across the electrodes emits an amount of RFI (radio
frequency interference) which can cause problems with the measuring electronics
and can possibly, over long exposure, have an affect on the human body when
in close proximity. This is part of the reason for the Lead Mesh shielding
built into the sample stand door and the RF shielding around the door frame
of the MOA.
- As the sample is being analyzed, a fire hazard is possible by the burning
sample. Generally analyzing a sample will not cause problems, but certain
samples (especially power plant fuels when used with the MOA GT) can catch
fire. In the event of “flaming” of the sample, the closed door will limit
any possible damage, to the interior of the MOA sample stand. For more information
please see Antares Technical Bulletin of April 2002: Fire Hazards in MOA
/ FAS Sample Stand.
Source Interlock Operational Description: This interlock system
“turns on” the MOA source by providing a safety circuit the 110 VAC to power
the various source functions needed to spark the oil / fuel sample in the MOA
sample stand.
- The Interlock system uses the 110 VAC generated by the closing of the K2
“source on” relay. This K2 “source on” relay is located on a small circuit
board, on the floor of the MOA electronics cabinet and should not be confused
with K1 / K2 relay located in the source chassis. This K2 relay is closed
by a signal from the MOA control electronics, when the burn switch is pushed.
- Closing of the K2 relay allows 110 VAC to pass through the K2 closed contacts,
through the closed door interlock switch, and into the MOA source closing
the contacts of the K1 power relay.
- It is the K1 power relay which supplies power, through 3 sets of contacts,
to A) the blower fan inside the source, B) the disc electrode
motor inside the sample stand and C) power to the Pulse and Ignition
circuits.
- It is the Pulse & Ignition circuits which sparks the oil / fuel sample
at the analytical gap. If either the K1 Power Relay (source chassis) or the
K2 “source on” relay (located on the card on the electronics cabinet) fail
to close … the source will not fire.
Troubleshooting and Repair:
- The most common failure of the interlock system will be failure of the interlock
switch located on upper right corner of the MOA sample stand door frame.
This is an “ON-OFF-ON” switch which is always OFF when the door is open.
- When the door is closed the switch will be pushed into the ON position (the
switch is spring loaded and will always go into the OFF position when the
door is opened), the interlock circuit will be completed and the source will
burn when the Burn switch is pushed.
- The 2nd ON position is: when the stand door is opened the switch
button can be pulled out and should lock into place. In this position the
source should also fire. THIS SWITCH POSITION IS ONLY TO BE USED FOR TEST
PURPOSES. CONSTANT USE IN THIS SWITCH POSITION, BECAUSE THE SAMPLE STAND DOOR
WILL BE OPEN, WILL RESULT IN DAMAGE TO THE SPECTROMETER, AND / OR POSSIBLE
INJURY OR DEATH TO OPERATING PERSONNEL.
- After all appropriate safeguards are taken (no personnel near the sample
stand electrodes, eye protection GLASS ONLY … not plastic etc.) ….
with the door opened, disc & rod electrodes in the stand and NO
SAMPLE IN THE STAND.
- Pull the switch out until it “clicks.” The switch will stay open and the
interlock switch is now “ON.”
- WARNING: For the next step the source will start and the following will
happen A) the source will start a burn across the analytical gap and there
will be a BRIGHT ARC across the electrodes … DO NOT look at this light
directly unless wearing GLASS eye protection (eye protection from strong UV
light), B) The electrode motor will start to rotate the disc electrode
and C) The blower motor in the source will start to rotate. NOTE: “blower
motor” means the motor fan inside the source which blows air across the Auxiliary
Gap. This is not the exhaust motor located above the sample stand.
If 1 or more of these 3 things do not happen immediately hit the Stop /
Reset button and check the following:
- The switch is in the pulled out position.
- The main power in on the MOA and Ready / Standby switch is in the Ready
position
- Fuses F6, F7 and F10 are good. These fuses are located at the fuse panel
on the right side of the spectrometer.
If the source now fires: close the door and try to start the source
again. If the source fires with the door open and interlock switch in the pulled
out position but not with the door closed. This is a good indication you have
a bad switch. Replace the switch.
If, after checking all the above: the source still does not fire and / or the disc electrode
motor does not rotate then:
- Turn off all power to the spectrometer and remove the power cable.
- Using an ohm meter check for continuity of the switch contacts in both “ON”
positions. If no continuity is found, replace the switch.
- If continuity is found then the problem is somewhere else in the “Source
On” circuitry and that will be covered in the May Antares Analytical Tech
Bulletin.
____________________________________________________________________________
This Technical Bulletin is written
by Gerry Smith, Technical Services Manager of Antares Analytical. Having spent
22 years with Baird in various capacities, including many years as Final Test
/ Calibration Manager and Senior Service Engineer, International Operations,
he or any of our Technical Support Staff, are well qualified to answer any Technical
or Applications problems you may have.
For further information or access to past monthly Technical Bulletins covering
a variety of subjects, please visit our website (still under development) at
www.antaresanalytical.com .For
more information, to be added to this list, or help with other subjects including
Tech Support, spare & replacement parts, or on-site Service by qualified
Antares Analytical Service Engineers, please contact us at info@antaresanalytical.com
This is intended to give Baird spectrometers
users a better understanding of your Baird spectrometer and help you get best
results from your spectrometer at the lowest reasonable cost. To
be removed from our email list, please reply with REMOVE in the subject
heading and we will never bother you again unless you request.
This bulletin is meant for informational
purposes only and it not meant as a substitute for Service performed by a qualified
Service Engineer. Any use of this information is done strictly at the discretion
of the user. Antares Analytical is not responsible for any misfortune which
results as the result of use of these Technical Bulletins, in place of service
by a Qualified Service Engineer.